Key Features

All features of the 3000 in a rugged transport case

  • 0 – 1000V DC / AC Voltage
  • 0 – 30A DC / AC Current
  • 0 – 1 GOhm Resistance
  • 1nF – 10mF Capacitance
  • 1Hz – 10MHz Frequency
  • Range of external accessories to increase workload

Description

The 9000A series range of multiproduct calibrators offers the widest workload of any calibration system presently available. The 9000A series provides all of the functionality of the 3000A series multiproduct calibrator in a dedicated, rugged case for the ultimate in on-site calibration performance.

Using an innovative series of internal and external options, the 9000A provides workload coverage that no other manufacturer offers. With external options ranging from Thermocouple source and measurement to a dedicated low current source, and internal options providing Oscilloscope calibration to 600MHz and calibration of power meters with power quality functionality.

The 9000A series is designed for use in industrial environments including offshore, aerospace and military applications. The rugged case with integrated wheels and retractable handle lends itself to easy transportation to and from site. A fast warm-up time ensures calibration can be performed quickly and accurately.

WIDE WORKLOAD COVERAGE

The 9000A series offers the widest workload coverage of any multi-product calibrator, with the ability to calibrate 18 types of instrument. It has the power to drive older analogue instruments alongside the accuracy to calibrate modern digital instruments.

COST-EFFECTIVE CALIBRATION

The 9000A series features three models designed to suit every budget and accuracy requirement, and with a wide range of options that allow your calibrator to be tailored to meet your needs, the 3000 series is expandable for future needs.

RUGGED PORTABILITY

A dedicated transportable design that is ideal for industrial use. A fast warm-up time combined with the latest technology, all included in a rugged and portable design.

 

SOFTWARE COMPATIBILITY

As with all Transmille products, compatibility with ProCal calibration software is guaranteed. Existing procedures will not need to be re-written, due to Transmille’s unique implementation of flexible standards support where procedures are written around the UUT, not the standard.

All Models

Choose the perfect Model

9050A Model

9041A Model

DC Voltage
Best 1 Year Total Accuracy
0 – ±1000V
±50 ppm
0 – ±1000V
±25 ppm
DC Current
Best 1 Year Total Accuracy
0 – ±22A
±100 ppm
0 – ±30A
±50 ppm
AC Voltage
Best 1 Year Total Accuracy
20mV – 1000V, 10Hz - 100kHz
±0.035%
20mV – 1000V, 10Hz - 500kHz
± 0.035%
AC Current
Best 1 Year Total Accuracy
20uA – 22A, 10Hz – 10kHz
±0.07%
20uA – 30A, 10Hz – 10kHz
±0.06%
Resistance (Passive Decade)
Best 1 Year Total Accuracy
0 Ohms – 100 Ohms
±60 ppm
0 Ohms – 1 GOhm
±40 ppm
Capacitance (Passive Decade)
Best 1 Year Total Accuracy
10nF - 1uF
±0.25%
10nF - 10uF
±0.25%
Frequency
Best 1 Year Total Accuracy
1Hz – 10MHz
±20 ppm
1Hz – 10MHz
±20 ppm (1 ppm option)
Temperature (Thermocouple)
Best 1 Year Total Accuracy
J,K,T,R,S,B,U,N,E,L,U,C
±0.09°C
J,K,T,R,S,B,U,N,E,L,U,C
±0.09°C
Dry Block Temperature (Option)
1 Year Total Accuracy
25°C - 200°C
±1°C
25°C - 200°C
±1°C
Black Body Source Temperature (Option)
1 Year Total Accuracy
25°C - 200°C
±1°C
25°C - 200°C
±1°C

Internal Options (Choose At Time Of Order)

PRT/RTD Temperature Simulation (Option PRT, Included in -SYS Models)

The PRT option simplifies calibration of high-accuracy resistance thermometers. Reliable readings are guaranteed regardless of measurement technique, because the calibrator uses passive precision resistors.When the PRT option is calibrated, the precise temperature values on the ITS-90 scale for PT-385 resistors are stored in the memory.  The values can be recalled and displayed every time a specific resistor is chosen, to allow accurate calibration.

When combined with the SIMRC option, an additional 2-Wire variable output is available, which enables simulation of any value.

Simulated Resistance & Capacitance (Option SIMRC, Included in -SYS Models)

This option allows for resistance calibration between cardinal points. It has a continuous resistance range between 0 Ohms and 1 GOhms, with variable capacitance from 0.95uF to 100mF.

This option enables calibration of multimeters with non-decade calibration points.

Inductance Calibration Option (IND)

Induction

Oscilloscope Calibration (Option SCP)

Advanced oscilloscope calibration option, adding the ability to calibrate oscilloscopes to your multiproduct calibrator.

Waveforms for amplitude, timebase, rise time, and bandwidth are provided, with outputs up to 300V pk-pk.

SCP250 enables calibration of oscilloscopes up to 250 MHz.

SCP350 enables calibration of oscilloscopes up to 350 MHz.

SCP600 enables calibration of oscilloscopes up to 600 MHz.

(Option PWRSINE, included in -SYS Model)

The PWRSINE option enables cost-effective calibration of power (watts) and VA ranges on power meters and clamp meters with power measurement functions. The PWRSINE option provides variable voltage and current simultaneously, with an accurately adjustable phase relationship.

Any voltage up to 1025V can be set, with a maximum current output of 30A without the need for external current amplifiers, allowing up to 1.5MW of power to be simulated.

The power function is very easy to use through intuitive menu design. Simply select ‘Power’ from the calibrator menu, enter the required parameters and connect the power meter to the indicated terminals. Voltage, current, frequency and phase can all be entered using the numerical keyboard, with phase angle adjustable to a resolution of 0.1°.

Power output is automatically calculated, and power factor and apparent power are also displayed on the screen automatically for quick comparison.

(Option PWRDDS)

The PWRDDS option enables cost-effective calibration of power (watts) and VA ranges on power meters and clamp meters with power measurement functions. The PWRDDS option provides variable voltage and current simultaneously with an accurately adjustable phase relationship. In addition to the functions provided by PWRSINE, PWRDDS enables harmonics and distortion to be added to the power output for calibration of power quality functions of advanced power meters.

Pre-programmed harmonics are loaded into memory, with custom waveform shapes able to be loaded into the calibrator through the included computer software for completely customized testing with waveforms with custom harmonic content.

 

Any voltage up to 1025V can be set, with a maximum current output of 30A without the need for external current amplifiers, allowing up to 1.5MW of power to be simulated.

The power function is very easy to use through intuitive menu design. Simply select ‘Power’ from the calibrator menu, enter the required parameters and connect the power meter to the indicated terminals. Voltage, current, frequency and phase can all be entered using the numerical keyboard, with phase angle adjustable to a resolution of 0.1°.

Power output is automatically calculated, and power factor and apparent power are also displayed on the screen automatically for quick comparison.

External Adaptors and Accessories (Add Anytime)

Thermocouple Simulation & Measurement Adapter (EA001A, Included in -SYS Models)

For the ultimate in accuracy, Transmille has designed a dedicated source/measure external adapter. This keeps the electronics required to generate the low-level signals used for thermocouple simulation as close as possible to the measuring input of the thermometer, which also allows the signal to be as free from electrical noise as possible, and eliminates errors caused by heat from surrounding electronics.

The cold junction compensation (CJC) sensor is mounted in the thermocouple plug itself. By measuring the cold junction at the instrument’s input, you can simulate any type of thermocouple without using thermocouple compensation cables.

Features

  • Enables fast and accurate calibration of thermocouple meters
  • Uses an innovative external measurement of cold junction temperature for ultimate accuracy
  • Utilizing a thermocouple plug removes the requirement for expensive thermocouple compensation cable

2/10/50 Turn Coil Adapter (EA002)

The Transmille EA002 offers several unique features.

Robust construction allows for years of hard use without risking damage to the coil.  An included alignment table ensures repeatable results.  This option was designed for the calibration of wound AC and magnetic field hall effect AC/DC clamp meters.

Features

  • Calibrates clamp meters up to 1500A (30A input)
  • Wide range of clamp sizes covered
  • 2/10/50 Turn Coils
  • High accuracy AC/DC design, with minimal errors due to alignment
  • Alignment table included for accurate and repeatable readings

Optical Tachometer Calibration Adapter (EA003)

Optical adapters can be calibrated quickly and easily, thanks to a high-intensity LED light source. Using the keypad, or ProCal software for automated calibration, simply enter the RPM value. With a range of 60 – 60,000 RPM, the EA003 can calibrate a whole range of optical tachometers.

Features

  • High-intensity LED light source
  • Direct keyboard input in RPM
  • Control via ProCal for fully automated calibration
  • Output range from 60 to 60,000 RPM

High Resistance/pA Measurement Adapter (EA008)

This option eliminates the need for a separate electrometer. The low-cost EA008 provides five current measurement ranges, down to 10nA, with a resolution of 1pA. It also incorporates a ‘virtual ground’ input, avoiding input impedance errors.

The EA008 allows high-accuracy resistance measurements at any voltage up to 1000V, with the high voltage output from the 4000 Series calibrators.

Features

  • Accurate current measurement down to 1pA resolution
  • Ideal for high value resistance measurements up to 1 TOhm
  • High resistance measurement voltages up to 1000V
  • 5 current measurement rnages from 10nA to 100uA
  • Control directly from the calibrator front panel or use with ProCal software

100A DC/70A AC Transconductance Amplifier (EA3012A)

The EA3012A is a powerful linear transconductance amplifier designed for use with Transmille calibrators to provide a cost-effective solution for generation of high currents.

The EA3012A connects directly to the adapter input connector on the front of Transmille calibrators and is controlled directly via the front panel with output in terms of Amps, or can be automated through ProCal calibration software.

Higher currents are available by paralleling multiple EA3012A amplifiers to reach currents in excess of 300A.

Features

  • 100 DC/70A AC up to 400Hz
  • Stackable for higher currents
  • Low-noise ultra linear MOSFET technology
  • Controlled directly via front panel of calibrator
  • Automation through ProCal software
  • Ideal for calibrating high current shunts, meters and clamp meters

High Resistance/pA Measurement Adapter (EA013)

This option eliminates the need for a separate electrometer. The low-cost EA008 provides five current measurement ranges, down to 10nA, with a resolution of 1pA. It also incorporates a ‘virtual ground’ input, avoiding input impedance errors.

The EA008 allows high-accuracy resistance measurements at any voltage up to 1000V, with the high voltage output from the 4000 Series calibrators.

Features

  • Ideal for calibration of electrometer instruments
  • Source accurate current down to picoamp levels
  • Low open circuit compliance voltage
  • Safe to use on sensitive input circuits
  • Controlled directly from the calibrator front panel or via supplied software
  • Compatible with ProCal for full measurement automation

Torque Calibration Adapter (EA014)

Connecting the EA014 to your Transmille calibrator adds torque calibration capabilities to your laboratory in a cost-effective way.

An external torque transducer allows you to calibrate torque screwdrivers up to 10Nm, with a range of torque transducers from 1Nm to 1500Nm.

Multi-Function Workstation (EA015)

The Multi-Function Workstation is a central work area, where all connections are conveniently routed to terminals mounted on the adapter front.

This workstation provides a versatile platform for efficient calibration of equipment, incorporating the advanced clamp coil, thermocouple simulation and tachometer calibration adapters.  It also includes process control source/measure capabilities.

Features

  • Insulation Testers
  • Continuity Testers
  • Optical Tachometers
  • Thermocouple Meters
  • Clamp Meters
  • Source functions of process control meters

Temperature & Humidity Monitoring Adapter (EA016)

The EA016 allows measurement of the local temperature and humidity for recording on certificates and test reports.

Working on-site makes it essential to maintain accurate calibration records.  The EA016 avoids the requirement of humidity/temperature probes.

Features

  • Humidity measurement from 10% to 90%
  • Temperature measurement from 0°C to 50°C

Option Documentation

EA016-Example-Certificate

4-Channel Oscilloscope Test Head Adapter (EA017)

The EA017 provides a 4-channel switching test head for automation of oscilloscope calibration.

You can easily select a low impedance buffered amplitude output for accurate calibration of 50 Ohm output oscilloscopes.

Features

  • 4 Output channels for automating oscilloscope calibration
  • Minimizes lead changing when calibrating oscilloscopes
  • Choose between direct or buffered outputs for amplitude calibration

Option Documentation

EA017-Operation-Guide—web

3000 Lead Set (3000LEAD)

Provided in a hard carry case for storage, the 3000 Leadset provides a comprehensive set of leads for making connections to typical test equipment.

These leads and connection accessories have been carefully constructed to minimize lead and connection errors and ensure safety. Non-retractable shrouds on voltage leads eliminate the risk of accidental shocks due to non-touch-proof leads.

3A/63V Power Supply Calibration Adapter (EA3023)

The EA3023 is an essential option for simplifying calibration of power supplies. The PSU adapter measures the output voltage, and provides a precision electronic load. This allows volts, current, and even output resistance to be calibrated in one compact unit. Accuracy and dependability are assured through 4-wire connection.

The output resistance of the PSU can be easily measured by measuring the on load/off load voltages of a power supply. Operation is controlled through the virtual front panel software (supplied) or ProCal.

Features

  • Measures voltage up to 63V
  • Active current load to 3A
  • Calculates output resistance of a power supply for complete calibration

AC/DC Kilovolt Amplifier (EA3024)

The EA3024, with its advanced electronics, is a simple and safe, cost-effective solution for calibration of high voltage probes and dividers.

It can be controlled through dedicated front panel software, or ProCal for safe calibration at high voltages.

Features

  • Outputs from 1kV to 10kV DC
  • Outputs from 1kV to 5kV AC at up to 60Hz
  • Operated using supplied software or via ProCal calibration software

60A/60V Power Supply Calibration Adapter (EA3025A)

The EA3025A is an easy-to-use, low-cost solution for calibrating high current power supplies.

An accurate, stable electronic load simulates current load on power supplies, and allows measurement of voltage and output resistance.  It minimizes equipment required for calibrating these common devices.

Features

  • Easy to use, low cost solution
  • Accurate, stable electronic load
  • Minimizes equipment required for calibration

TPMXX Pressure Modules (TPM001 to TPM018)

A modern calibration laboratory must be able to perform calibration in multiple disciplines in order to remain competitive. Transmille has a range of pressure modules to assist, adding pressure management capability to our products.

Choose from a range of pressure modules that connect directly to our products, and provide read-back directly on-screen.